Table 5. Percent differences between MicroCell counts and hemacytometer counts, by technician, for both SFF study counts and QC counts*
    Difference Between MicroCell Counts and Hemacytometer Counts
  Technician   SFF Study Data (%)   QC Data (%)

  Standard 1   NA   38
  Standard 2   NA   26
  Technician 1   28   40
  Technician 2   13   19
  Technician 3   32   29
  Technician 4   11   33
  Technician 5   36   20
  Mean % difference for all samples, study site   24   29
     technicians only   n = 509 study subjects{dagger}   n = 128 replicates from 32 ejaculates

  *QC indicates quality control; SFF, Study for Future Families
  {dagger}The SFF study data are given in Brazil et al, 2004